Technology

Know-how and calibration

We assemble and calibrate our sensors in-house. Every sensor is qualified for sensitivity and frequency response.

Know-how

Our sensors are assembled on our premises. We work with different optical fibers and coatings, using telecom components wherever possible because they are both high-performing and affordable.

The membranes are made by an industrial specialist in clean-room microfabrication, which ensures the repeatability and consistency of our sensors. Made of silicon or silicon nitride, they offer high stability and good long-term chemical resistance.

At the heart of our microphones, ceramic or borosilicate ferrules provide good chemical and mechanical resistance. We master many bonding techniques, which lets us choose the best one for each type of sensor and customer need.

Our tests show that our sensors withstand very high temperatures without any change in how they work. And because every environment is different, we adapt our sensors on request: size, membrane thickness, pressure range.

Calibration

All our sensors are calibrated in-house. We qualify their sensitivity and frequency response, as well as the influence of temperature, ambient pressure and humidity, with the following equipment:

  • B&K type 4231 acoustic calibratorSensitivity calibration at 1kHz at 94dB and 114dB
  • Acoustic speaker in anechoic box: B&K type 4232Free-Field calibration of sensitivity and frequency response between 100Hz and 10kHz.
  • GRAS RA0014 electrostatic actuator accompanied by its 14AA amplifier.Pressure-Field calibration of the frequency response between 100Hz and 200kHz
  • IKA 125 control ovenTemperature test up to 300°C
  • Carbolite CWF1200 ovenTemperature test up to 1200°C
  • Vacuum bell with sealed optical and electrical crossingUp to 100mBar
  • Overpressure box with sealed crossingsUp to 5 bars
  • Aaronia Spectran NF-5030 X Spectrum AnalyzerFrom 1Hz to 1MHz
  • Dewesoft acquisition system: SiriusDAQ 8 channels 16bit 1MHz simultaneous
  • Michelson fiber interferometer

We therefore know the exact response of each sensor and how external conditions affect it, as the examples below show.

A challenging measurement environment?

Tell us about your application: temperature, pressure, radiation, distance, installation constraints. We will propose the right sensor and fiber.

Discuss your project Technology